Skip to main content

Table 4 Secondary endpoints—freeze parameters

From: Comparison of pulmonary vein isolation between two commercially available cryoballoon systems

 

Minimal temperature (°C, median, interquartile range)

Temperature at TTI (°C, median, interquartile range)

TTI (s, median, interquartile range)

Arctic Front™ (n = 306)

PolarX™ (n = 146)

p value

Arctic Front™ (n = 306)

PolarX™ (n = 146)

p value

Arctic Front (n = 306)

PolarX™ (n = 146)

p value

LSPV

− 52 (− 55 to − 47)

− 56.5 (− 60 to − 55)

< 0.001

− 40 (− 45 to − 33)

− 48.5 (− 50 to − 44)

< 0.001

49 (37–64)

60 (38–74)

0.819

LIPV

− 48 (− 52 to − 47)

− 53 (− 56 to − 51)

< 0.001

− 37 (− 42 to − 27)

− 42.5 (− 47 to − 40)

< 0.001

39 (25–74)

56.5 (35–87)

0.218

RSPV

− 54 (− 57 to − 51)

− 60 (− 64 to − 53)

< 0.001

− 39.5 (− 43 to − 32)

− 42.5 (− 50 to − 35.0)

< 0.001

40(28–66)

44.5 (25–79)

0.424

RIPV

− 51 (− 54 to − 46)

− 57 (− 65 to − 53)

< 0.001

− 40 (− 42 to − 35)

− 43 (− 49 to − 34)

< 0.001

52.5 (36–90)

47.5 (31–100)

0.062

  1. Bold = p<0.05 (statistically significant)
  2. TTI, time to isolation; LSPV, left superior pulmonary vein; LIPV, left inferior pulmonary vein; RSPV, right superior pulmonary vein; RIPV, right inferior pulmonary vein